Biosafety Cabinet Class II B2 LBS2-B22 is a Class II type B2 total exhaust cabinet with 420 mm (17 inch) maximum opening. It has ULPA Filter Two in number, 99.999 % efficient at 0.12 µm and offers two layer laminated toughened glass >5 mm. Designed with Large digital LCD display used to monitor all safety parameters and Universal caster for moving with brake and levelling feet. Work area surrounded by negative pressure and it can ensure maximum safety in work area.
Max Opening
500 mm (20 inch)
Tested Opening
Safety Height 200 mm(8 inch)
Inflow Velocity
0.53 ± 0.025 m/s
Down flow Velocity
0.33 ± 0.025 m/s
Pre-Filter
Washable
ULPA Filter
99.999 % efficiency at 0.12 μm, filter life indicator
Front Window
Motorized, two-layer laminated toughened glass ≥ 5 mm, anti UV
Noise
NSF49 ≤ 61 dB / EN1246949 ≤ 58 dB
Illumination
≥ 1000 Lux
Consumption
900 W
Display
LCD display, exhaust filter and down flow filter pressure, filter and UV lamp working time, inflow and down flow velocity, filter life, humidity and temperature, system working time etc.
Waterproof Sockets
Two, total load of two sockets 500 W
Control System
Microprocessor
Airflow System
0 % air recirculation, 100 % air exhaust
Alarm
Abnormal airflow velocity, Filter replacement, Front window at unsafe height
Exhaust Duct
4 meters PVC duct, Diameter 300 mm
Material Work Zone
304 stainless steel
Material Main Body
Cold-rolled steel with anti-bacteria powder coating
Work Surface Height
750 mm
Caster
Foot master caster
UV Lamp
30 W × 1 UV timer, UV life indicator, emission of 253.7 nano meters for most efficient decontamination
Illuminating Lamp
LED Lamp
16 W × 2
Internal dimension (W×D×H)
1350 × 600 × 660 mm
External dimension (W×D×H)
1500 × 760 × 2250 mm
Power supply
AC 220V±10 %, 50/60 Hz; 110V±10 %, 60 Hz
Weight
302 kg
Designed with ULPA filter and UV indicator
Automatic air speed adjustable with filter block
Large digital display used to monitor exhaust filter and down flow filter pressure, filter and UV lamp working time, inflow and down flow velocity, filter life, humidity and temperature, system working time etc
Two layer laminated toughened glass >5 mm, anti UV
Universal caster with brake and levelling feet
Memory function in case of power-failure
Equipped with Interlock function, UV lamp, front window, fluorescent lamp
Designed with Remote control and Audio, visual alarm
Time reserve function
Motorized front window
Maximum safety in work area surrounded by negative pressure
Adjust front window height by foot during experiment, to avoid airflow turbulence caused by arm movement
Biosafety Cabinet product is used in protection from biological hazards and contamination as well as protection from volatile toxic chemicals.