Analytical HPLC-A7832
Detector Specifications
Model No HPD-A20 HPD-A21
Wavelength range Single wavelength ( 190 to 800 nm ) Dual wavelength ( 190 to 800 nm )
Wavelength accuracy ± 0.5 nm ± 2 nm
Wavelength repeatability 0.15 nm 0.3 nm
Band width 8 nm 8 nm
Optical path 10 nm 10 nm
Deuterium lamp 190 to 400 nm 190 to 400 nm
Tungsten lamp 400 to 800 nm 400 to 800 nm
Baseline noise ( Static ) Au ≤ ± 0.35 × 10-5 ≤ ± 0.5 × 10-5
Baseline drift ( Static ) Au / h ≤ 1 × 10-4 ≤ 1.5 × 10-4
Minimum detection limit ≤ 4 × 10-8 ≤ 4 × 10-8
Power 100 W 100 W
Power supply 90 V / 50 to 60 Hz 90 V / 50 to 60 Hz
Dimension 460 × 315 × 180 mm 460 × 315 × 180 mm
Weight 9.5 kg 9.5 kg
Pump Specifications Flow rate rangeMaximum pressureColumn sizeSample sizePressure pulsationFlow rate accuracyFlow rate repeatability %Gradient accuracy %Gradient repeatability %DimensionPower supplyWeight
0.0001 to 10 ml / min
50 MPa , 0 to 2 ml / min
4.6 × 250 mm
5 µm
≤ 1 % at 10 MPa
± 0.15 % at 1 ml / min
≤ 0.1
± 1.5
≤ 0.2
460 × 310 × 180 mm
90 V
11.6 kg
Analytical HPLC-A7835
Pump Specifications Model NoFlow rate rangeMaximum pressureColumn sizeSample sizePressure pulsationFlow rate accuracyFlow rate repeatability %Gradient accuracy %Gradient repeatability %DimensionPower supplyWeight
HPP-B1
0.0001 to 10 ml / min
42 MPa , 0 to 5 ml / min
4.6 × 250 mm
5 μm
≤ 1.5 % at 10 MPa
± 0.2 % at 1 ml / min
≤ 0.15
± 2
≤ 0.2
450 × 260 × 190 mm
90 V to 240 V
11.2 kg
Detector Specifications Model NoWavelength rangeWavelength accuracyWavelength repeatabilityBand widthOptical pathDeuterium lampTungsten lampBaseline noise ( Static ) AuBaseline drift ( Static ) Au / hMinimum detection limitPowerPower supplyDimensionWeight
HPD-B2
190 to 800 nm
± 0.75 nm
0.2 nm
8 nm
10 nm
190 to 400 nm
400 to 800 nm
≤ ± 0.5 × 10 -5
≤ 1.5 x 10 -4
≤ 4 x 10 -8
100 W
90 V / 50 to 60 Hz
460 × 260 × 190 mm
9.2 kg
Analytical HPLC-A7840
Pump Specifications Solvent deliveryPump HeadFlow rate rangeFlow rate incrementMaximum pressureColumn sizePipeline SizeFlow rate accuracyFlow rate precisionGradient accuracy %Gradient repeatability %GLPDisplayPowerPower supplyDimension
Double plunger series mode, floating plunger design, single stroke volume 30 µl
10 ml stainless steel pump head
0.001 to 9.999 ml / min
0.001 ml/min
42 MPa
4.6 × 250 mm 5 µm
1/16 inch
± 0.5 %
RSD ≤ 0.1%
± 0.5 %
RSD ≤ 0.15%
Automatic recording of pump sealing ring working time
LCD backlight display
80 W
220 V ± 10%, 50 Hz
398 × 149 × 267 mm
Detector Specifications Wavelength rangeWavelength accuracyWavelength repeatabilityDeuterium lampTungsten lampBaseline noise ( Static ) AuBaseline drift ( Static ) Au / hTime constantsDisplayGLPPowerPower supplyDimension (L × W × H)
190 to 700 nm
± 1 nm
0.2 nm
190 to 400 nm
400 to 700 nm
1 × 10-5 AU/h at 254 nm with methanol/water= 80/20
1.5 × 10-4 AU/h at 254 nm with methanol/ water= 80/20
16 grades adjustable
LCD backlight display
Total number of lights, lamp replacement time and product serial number
75 W
220 V ± 10%, 50 Hz
398 × 149 × 267 mm
Analytical HPLC System LAHS-A10
Pump specification TypeFlow Rate RangeFlow AccuracyFlow PrecisionMaximum PressurePressure accuracyPressure pulsationComposition rangComposition precision
Quaternary Low Pressure Pump
0.001 to 10.000 ml/min, in 0.001ml increments
≤ ± 0.2% (1.00 ml/min, H2 O, R.T., 8.5 MPa)
RSD ≤ 0.06% (1.00 ml/min, H2 O, R.T., 8.5MPa)
63 MPa
≤ ± 3.0 %
≤ 0.1 MPa
0 to 100 %
≤ 0.3 % RSD
UV/VIS Detector Wavelength RangeWavelength RepeatabilityWavelength AccuracyBand WidthResponse TimeBaseline NoiseDriftPath lengthLight Source
190 to 800 nm
≤ 0.1 nm
≤ ± 1.0 nm
6 nm
0.0 s to 5 s
≤ ± 0.25 × 10-5 AU(empty cell, 254 nm, 1.0s)
≤ 0.5 × 10-4 AU/h (empty cell, 254 nm, 60 min)
10 mm(analytical cell)
Deuterium Lamp and Tungsten Lamp
Data-processing Workstation System contain the tables for regular samples
Equipped with Internal and external standard calculation process
Automatic display, print, export of data and starting of another program on completion on measurement
Observe with the condition of 21 CFR part 11 directive of the FDA
System Kits 1 Set
Combined A/D converters
Combined vacuum degasser and 4 channel
Power Supply AC 110 /220 V ±1 0 %, 50/60 Hz
Analytical HPLC System LAHS-B10
High Pressure Constant Flow Pump Double plunger reciprocating pumpFlow rangeSetting step sizeFlow accuracyFlow stabilityMaximum pressureDisplay pressure errorPressure fluctuationPower consumption
1 set
0.001 to 10.00 mL/min
0.001 mL/min
≤ ± 0.3 %
RSD ≤ 00.1 %
42 MPa
≤ 1.0 %
≤ 1.0 %
80 W
UV/Visible Detector Wavelength rangeWavelength repeatabilityBand widthResponse timeLinearity rangeBaseline NoiseBaseline driftMinimum detectable concentrationDetector cellLight sourcePower consumptionVoltagePacking Dimensions (W × D × H)Gross weight
190 to 700 nm
± 0.1 nm
8 nm
0 to 4.9 seconds
1.8 AU (5 %)
≤ ± 1 × 10-5AU
≤ 2.0 × 10-4AU/h
4.0 × 10-9 g/mL
Optical path: 10 mm
Volume: 10 µL
Deuterium lamp
100 W
AC 110 V/ 220 V ± 10 %, 50/60 Hz
600 × 440 × 380 mm
2 × (600 × 410 × 300 mm
54 kg
Analytical HPLC System LAHS-B11
High Pressure Constant Flow Pump Double plunger reciprocating pumpFlow rangeSetting step sizeFlow accuracyFlow stabilityMaximum pressureDisplay pressure errorPressure fluctuationPower consumption
2 sets
0.001 to 10.00 mL/min
0.001 mL/min
≤ ± 0.3 %
RSD ≤ 00.1 %
42 MPa
≤ 1.0 %
≤ 1.0 %
80 W
UV/Visible Detector Wavelength rangeWavelength repeatabilityBand widthResponse timeLinearity rangeBaseline NoiseBaseline driftMinimum detectable concentrationDetector cellLight sourcePower consumptionVoltagePacking Dimensions (W × D × H)Gross weight
190 to 700 nm
± 0.1 nm
8 nm
0 to 4.9 seconds
1.8 AU (5 %)
≤ ± 1 × 10-5AU
≤ 2.0 × 10-4AU/h
4.0 × 10-9 g/mL
Optical path: 10 mm
Volume: 10 µL
Deuterium lamp
100 W
AC 110 V/ 220 V ± 10 %, 50/60 Hz
600 × 440 × 380 mm
3 × (600 × 410 × 300 mm
75 kg